• To avoid resonance, the natural frequency of a system must differ from the input frequency
  • When considering the input frequency it is essential to understand that any non-sinusoidal wave can be deconstructed into a series of harmonics through Fourier’s analysis:
    • Each harmonic has increasing frequency and decreasing amplitude
Graph of amokitude against frequency demonstrating resonance at a the natural frequency of a system
  • In general, to avoid resonance, it is vital to ensure that the natural frequency of a system is greater than the 8th harmonic of the input system:
    • An invasive arterial blood pressure trace is assumed to have a maximum fundamental frequency of 3 Hz (equivalent to a heart rate of 180 bpm)
    • The natural frequency of the measurement system must therefore have a natural frequency of >24Hz (above the frequency of the 8th harmonic) to avoid resonance
  • The equation below determines the natural frequency of a system, where k is stiffness and m is mass.
Graph of amokitude against frequency demonstrating resonance at a the natural frequency of a system
  • In practice, using invasive blood pressure monitoring as an example, the natural frequency of a system can be manipulated by changing the
    • Stiffness of the system: radius of tubing, elasticity of tubing material
    • Mass of the system: length of tubing, density of fluid